7 Guideline value for power settings
When performing the power settings, the user should generally
take into account the cutting speed, geometry of the electrode
Monopolar cutting:
Smooth cuts with thin needle electrode
Cutting with knife electrode, lancet or wire
loop
Cutting in wet environment (TUR)
Monopolar coagulation:
Coagulation of large surfaces
Coagulation of small surfaces
Bipolar application:
Coagulation
Cutting
Cutting in water environment (TUR)
!
48
and properties of the tissue.
The recommended settings serve as guideline values only and
must be adjusted to the relevant situation.
In the event of unfavourable conditions, these values may
Recommended settings
5 - 40 W
50 - 175 W
80 - 150 W
Recommended settings
60 W
40 W
Recommended settings
10 - 30 W
40 - 60 W
50 - 75 W
deviate!