Revert to factory default settings
The continually displayed CAL symbol refers to a previously
made adjustment.
What to do
1.
Bring up the CAL
mode
2.
Deactivate CAL
mode
3.
Revert to factory
setting
Important:
When the Transferpette
offset is performed, which means that the volume is changed
across the entire volume range of the pipette by the same
amount. It is recommend that the adjustment be performed
at 50% of the nominal volume.
Note:
The instrument is permanently adjusted for watery solutions, but
it can also be set for solutions with varying density, viscosity and
temperature. The Transferpette
in every mode, with the exception of the GEL mode.
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
'on'. 'CAL' blinks.
Press one of the arrow
keys to deactivate the CAL
mode. The display changes
from 'on' to 'off'. 'CAL'
continues to blink.
Press the ENTER key. The
CAL symbol disappears.
The instrument has now
been reverted to factory
default setting.
®
-8/-12 electronic is adjusted, a volume
®
-8/-12 electronic can be adjusted
Easy Calibration
Keys to press
Display readout
1x
>3s
1x
TM
5 7