Disassembling The Manifold - BRAND Transferpette-8 electronic Instruccionesde Manejo

Ocultar thumbs Ver también para Transferpette-8 electronic:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Disassembling the Manifold

Removing of nose cones and seals for cleaning or replacing
1.
Unscrew the manifold from
the handle.
2.
Turn both closures of the
manifold housing cover 90°
and slide off the housing.
3.
Push the mounting tool on
the nose cone and unscrew
it (Fig. A and B).
4.
Push the piston unit down to
the bottom. Remove the seal
and inspect it and clean or
replace if necessary (Fig. C).
Note:
The seal (Abb. C) will either
remain inside the nose cone
or will stay on the piston
after the nose cone is
removed.
6.
If required, lubricate the
piston lightly with the supplied
silicone oil.
7.
Push the seal on the piston
with its flat side facing up-
ward. Use the mounting tool
to mount and tighten the
cleaned or new nose cone
(Fig. D).
8.
Reassemble the manifold
and check the instrument
for tightness (s. page 59
„Servicing").
6 0
A
Push the mounting tool on and unscrew the single nose cone
C
Remove the seal for cleaning
or replacement
B
D
Insert the seal and screw in
the cleaned or new nose cone

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Transferpette-12 electronic

Tabla de contenido