Assembly - ProMinent CLE 3-mA Instrucciones De Servicio

Célula de medición de cloro para cloro libre
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5
Pouring
Electrolyte
IMPORTANT
Do not touch, damage or bring into contact with greasy substances the
white membrane or the electrodes on the electrode shaft. The sensor
will not, in such cases, work accurately.
Replace the membrane cap or send the sensor to ProMinent to have the
electrodes cleaned.
NOTE
Carry out the following actions over a washbasin.
Assembling
membrane cap
ProMinent
®

Assembly

Remove the red cap completely from the nozzle and cut the nozzle at the
marked position to open the nozzle canal.
Remove the membrane cap cover and unscrew the membrane cap from
the electrode shaft.
Rinse the membrane cap and the electrode with a little electrolyte.
Fill the membrane cap up to the rim with electrolyte.
Remove air bubbles by lightly tapping the membrane cap on an even
surface.
Place the electrode shaft upright onto the filled membrane cap and twist
until the thread bites.
Rotate the electrode shaft until the vent hole is pointing upwards.
Slowly screw in the membrane cap by hand up to the stop. Excess
electrolyte will seep out of the vent hole as you screw the parts together.
Rinse away the excess electrolyte from your fingers and from the sensor
under running water.
There should be no air left in the membrane cap/electrolyte. Repeat the
above steps if there is still air present.
Assembly
25