AC connection
Power: 100 to 240 V AC +10/-15%, 47 to 63 Hz
6. Anchor cable with clamp bar.
7. Close and latch power supply terminal protection cover.
8. Tighten cable gland.
Current output/HART (channel 1)
1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
4. Signal cable screen is folded back over outer sheath and grounded beneath cable clamp.
5. Connect wires to terminals using wiring tool.
5
4
Active current output
Active current output
6. Tighten cable gland.
SITRANS FCT030 transmitter
A5E03650800-02, 03/2012
C
Ca+
Cp-
C
U
0
Ca+
+
-
DC connection
Power: 24 to 90 V DC +20/-20%
6
5
Passive current output
Passive current output
Cp-
C
Cp-
C
+
-
Ca+
7