Commissioning Tests - Air Techniques ScanX Swift View Instrucciones De Montaje Y Uso

Ocultar thumbs Ver también para ScanX Swift View:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Installation
EN
US
Incisor
Premolar
Molar
Bite wing
Check and adjust the specific X-ray unit in accordance with the standard values
i

8.3 Commissioning tests

The required tests (eg acceptance test) must be done in accordance with local rules and regula-
tions
Find out which tests are required
i
Carry out testing in accordance with local rules and regulations
i
Electrical safety checks
Carry out the electrical safety check according to national law (e g in accordance with IEC 62353)
i
Document the results
i
22
OBJ_DOKU-0000002448-001-mehrere.indd 22
without X-ray field
limitation
70 kV
mGycm
2
008 s
73
011 s
100
014 s
128
014 s
128
DC emitter, 6 mA
Tube length 30 cm
X-ray field limitation
0.8 x 1.2 in
2
(20 x 30 mm
)
2
70 kV
mGycm
2
008 s
15
011 s
21
014 s
27
014 s
27
X-ray field limitation
1.2 x 1.6 in
2
(30 x 40 mm
)
2
70 kV
mGycm
2
008 s
31
011 s
42
014 s
54
014 s
54
2142100008L41 1708V002
24.08.2017 14:41:46