Troubleshooting
Fault
Horizontal, grey lines in
the X-ray image, extending
beyond the left and right
image edge
EN
US
X-ray image is stretched
lengthwise with bright,
horizontal stripes
X-ray image split vertically
into two halves
X-ray image with small
bright spots or clouding
Lamination of the PSP de-
taches at the edge
38
OBJ_DOKU-0000002448-001-mehrere.indd 38
Probable cause
Transport slipping
Wrong barrier envelope or PSP
used
Dirt in the laser slit (eg hair, dust)
Micro scratches on the PSP
Wrong retainer system used
PSP handled incorrectly
Solution
Clean the transport mechanism,
i
replace the transport belts if
necessary
Only use original accessories
i
Clean the laser slit
i
Replace the PSP
i
Only use original PSP and film
i
retainer systems
Use the PSP correctly
i
Comply with the operating
i
instructions of the PSP and film
retainer systems