5.7 Circuit continuity test
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform the continuity test:
set the rotary switch at
connect black test lead to COM terminal, and red test lead to
VΩCAP
contact the tips of test probes to the points of measurement,
read the measurement result on the display; the beep will be acti-
vated when resistance values are below approx. 50 Ω.
5.8 Diode test
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors. Do not test the diode un-
der voltage.
To perform the diode test:
set the rotary switch at
press MODE/VFD button, to display
connect black test lead to COM terminal, and red test lead to
VΩCAP
contact the tips of test probes to the diode. The red test probe
should contact the anode and the black should contact cathode,
read the test result on the display – the forward voltage is dis-
played.
For a typical silicon rectifier diode, it is approx. 0.7 V, and for
a germanium diode it is approx. 0.3 V
For LEDs with a low power, typical voltage value is in the
range of 1.2...5.0 V depending on the colour.