Professional True RMS Digital Multimeter with OLED display / English
To deactivatethe power OFF functionality press and hold the "MODE" button before powering
ON from the OFF setting. When auto power OFF is deactivated, the Auto Power Off icon will
not be visible in the display.
12. REL button
The RELATIVE function zeros out the reading on the display and stores it as a reference.
Subsequent readings will be displayed as the relative difference between the actual measurement and
the stored reference value.
1. Momentarily press the REL button. The "
the relative reading. Momentarily press the REL button again to return to normal operation.
2. Press and hold the REL button until the "LOG" indicator reappears can Manual or automatic
storage of data.
3. Press and hold the REL button until the "VIEW" indicator reappears can manually view the stored
data.
4. In the LOG mode, Momentarily press the REL button can manually record a data.
NOTE:The meter does not Autorange when the Relative mode is active. The display will read
OL if the difference exceeds the range. When this occurs, exit REL and use the RANGE button
to select a higher range. REL does not work on Frequency, Duty Cycle, or Temperature.
13. MAX/MIN/AVG Button
1. Momentarily press the MAX/MIN button to activate the MAX/MIN/Average mode.
"MAX" will appear on the LCD display and the meter will display and hold the highest
reading. The meter will update the reading when a higher "max" occurs.
2. Momentarily press the MAX/MIN button again to view the lowest reading. "MIN" will
appear on the LCD display and the meter will display and hold the lowest reading. The meter
will update the reading when a lower "min" occurs.
3. Momentarily press the MAX/MIN button once more to view the average reading.
4. "AVG" will appear on the LCD display and the meter will display the running average.
5. The meter will update the reading when the average value changes.
6. Press and hold the MAX/MIN button to end MAX/MIN/Average and return to normal
operation.
7. In the LOG mode, Momentarily press MAX/MIN button can Automatically record data.
NOTE:The meter does not Autorange when the MAX/MIN/AVG mode is active. The
display will read OL if the range is exceeded. When this occurs, exit MAX/MIN/AVG
and use the RANGE button to select a higher range. MAX/MIN/AVG does not work on
Frequency, Duty Cycle or Temperature.
10
" indicator will appear on the LCD display along with
Professional True RMS Digital Multimeter with OLED display / English
14. HOLD
Button
To freeze the reading on the LCD display, momentarily press the HOLD
indicator will be displayed while the reading is being held. Momentarily press the HOLD
button again to exit HOLD and return to normal operation.
To turn the backlight on, press and hold the HOLD
turn the backlight off, press and hold the HOLD
In the VIEW mode , momentarily press the HOLD
been recorded.
15. PEAK Button
The PEAK function is accessible when measuring AC Voltage or Current. It captures and displays the
highest positive peak and the highest negative peak of the AC waveform.
1. Momentarily press the PEAK button to view the highest positive peak. Peak MAX will appear on
the LCD display and meter will display and hold the highest reading.
The meter will update the reading when a higher positive peak occurs.
2. Momentarily press the PEAK button a second time to view the highest negative peak. Peak MIN
will appear on the LCD display and the meter will display and hold the highest reading. The meter
will update the reading when a higher negative peak occurs.
3. Momentarily press the PEAK button again to exit PEAK and return to normal operation.
4. IN the LOG mode,press the PEAK button, the time interval of the automatic storage record is
increased.
5. In time and date settings,press the PEAK button, the number of flickering will be increased.
16. Turn on Bluetooth
Leave the rotating function switch away from the OFF position. press and hold the MODE
until the "
" symbol appears on the LCD display. Then it can be paired and communicated with
APP .
button. The "HOLD"
button until the backlight turns on. To
button until the backlight turns off.
button to delete all the data that has
button