1.2 Specifications Piston burette TITRONIC
CE Zeichen:
Country of origin:
The following dissolvents/titration reagents are allowed to be used:
− All common titration solutions.
− As reagent water and all non-aggressive non-organic and organic fluids are allowed. If using combustible
fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the chemical industry.
− For fluids with higher viscosity (≥ 5 mm
speed can be adjusted.
− Fluids with viscosity over 20mm
Display:
Power supply:
RS-232-C Interface:
Data bits:
Stop bit:
Start bit:
Parity:
Baud rate:
Address:
RS-232-1
RS-232-2
USB Interface:
USB –Typ B
USB –Typ A („Host")
Stirrer:
As per the low voltage directive 2014/35/EU;
test basis EN 61 010-1: 2011-07 for laboratory devices
As per the EMC directive 2014/30/EU; test basis EN 61 326 Part 1: 2012
As per the EMC directive 2011/65/EU; test basis EN 50 581: 2013-02
FCC part 15B and ICES 003
Germany, Made in Germany
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
power supply 100-240 V; 50-60 Hz,
power input: 30 VA
Input: 12 Volt DC, 500 mA.
Use the Power supply TZ 1853, Type No.: FW 7362M/12 only!
RS-232-C interface separated, Daisy Chain function available.
adjustable, 7 or 8 Bit (default: 8 Bit)
adjustable, 1 or 2 Bit (default: 1 Bit)
static 1 Bit
adjustable: even / odd / none
adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
adjustable, (0 to 15, default: 01)
for computer, input Daisy Chain
devices of SI Analytics, titrator TitroLine
- Burettes TITRONIC
- Balances of the types Mettler, Sartorius, Kern, Ohaus
(for more, please contact SI Analytics)
- Exit Daisy Chain
1 x USB-type A and 1 x USB-type B
for connecting a PC
for connecting:
- USB keyboard
- USB printer
- USB manual controller,
- USB data media e.g. USB stick
Plug connection with integrated low-voltage supply (15 V ⎓) for the TM 50 Stirrer
- USB Hub to connect more USB devices
incorporated in the bottom of the casing of the TITRONIC
®
300
7750,
500, TITRONIC
110 plus, TITRONIC
(Release: 30. August 2018)
universal/300,
®
300 piston burette