Xylem SI Analytics TITRONIC 300 Manual De Instrucciones página 89

Burette a piston
Ocultar thumbs Ver también para SI Analytics TITRONIC 300:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

89
After the unit is unlocked, confirm with <OK>. The unit will move up the whole way.
Fig. 72
Pull off the dosing unit and attach the new one just like it was before. The two struts of UV protection must
comply with the mark on the housing (Fig.73).
Fig. 73
Accept by pushing <OK>/<ENTER>. In case you changed the size of the dosing unit, you can choose which size
you attached:
Fig. 74
If you want to change the reagents data, you can reset all reagent data.