Components Description; Standard Equipment; Certification - Telstar BiOptima Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

2.3.

COMPONENTS DESCRIPTION

(See Annexes A1&A2)
2.4.

STANDARD EQUIPMENT

Two inlet sockets 2A - 460W each located on either side of the work zone. Insulation level
IP44.
UV kit
Easy-to-work UV germicidal lamp with a timer function. It can only be enabled in the
Disinfection screen (Chapter 4.5 point 2.4) or by means of the UV light Auto Power on
screen (Chapter 4.5 point 3.5.1.2). For safety reasons, this button is electrically linked with
the light button, and it switches off if the front window is opened.
Average disinfection intensity over the work zone of 200 µW/cm
after a working life of 8000 hours.
WARNING
In the event of using this accessory, you should consider the normal precautions in the use of
these types of lamps. Their use is only suitable for previous external sterilisation of materials
introduced in the cabinet. So this lamp accessory should be used before the start of work in
the cabinet and it is essential to keep it turned off when the operator is in front of the working
area or working on it. The glasses provide protection from direct radiation but UV rays are
reflected on polished metal parts and can seriously damage operator's eyes. That is why the
glass front must be placed into the closed position to be able to switch on the U.V light, then
the fans will automatically disconnect.
2.5.

CERTIFICATION

Quality control and tests results certification carried out in line with the applicable standards
EN 12469 & DIN 12980. See Certificate inside the cabinet.
UM-BiOptima-EN-1907
2
with 20% depreciation
7 / 39

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Bioptima cyto

Tabla de contenido