Precision Water Bath Quick Start Guide
This quick start guide is intended for initial start up only. For all other procedures you must refer to the manual. Also, if any
of these steps are not clear refer to the manual before proceeding.
Filling
(Only approved fluids are Filtered/Single Distilled Water and De-ionized Water)
Ensure the reservoir drain port is closed and that all plumbing connections are securely plumbed or capped. Also ensure any residue is
thoroughly removed from the reservoir before filling.
To avoid spilling, place your samples/trays into the bath before filling.
Slowly fill the reservoir. When adding, point the opening of a container away from yourself.
Fill the reservoir with a minimum of one inch of water and a maximum level one inch lower than the tank upper surface.
Starting
Do not run the bath until fluid is added to the reservoir. Have extra fluid on hand. If the bath does not start refer to the manual.
• Place the circuit protector located on the rear of the bath to the I position.
• The screen will momentarily display
The set point is the desired reservoir fluid temperature.
49.9°C
50.0°C
and then the home screen appears displaying the set point and the audible alarm status.
50.0°C
Press
to start the bath and display the fluid temperature. If the reservoir fluid
temperature is below the set point the heater
will start.
Page 1 of 2