IMPACT 1640
3
4
9.6
REPLACING THE TRANSDUCER
The following procedure must only be performed
i
by a Titan Authorized Service Center.
1. Unplug the unit.
2. Loosen and remove the two filter assembly bolts (Fig. 23,
Item 1). Slide the filter assembly from the cart.
1
3. Loosen and remove the two motor shroud screws (Fig. 24,
Item 1). Remove the motor shroud (2).
4. Loosen and remove the two motor cover screws (3).
Remove the motor cover (4).
5. Disconnect the transducer wire from the motor controller
(5).
6. Pull the grommet (6) out of the mounting plate and slide
it up the shaft of the transducer (7) until it is clear of the
mounting plate.
1
2
7. Using a wrench, loosen and remove the transducer (7)
from the filter housing (8). Carefully thread the transducer
wire out through the mounting plate.
8. Slide the grommet (6) off of the old transducer (7) and
onto the new transducer.
9. Thread the new transducer wire through the mounting
plate and back to the motor controller (5).
10. Thread the new transducer (7) into the filter housing (8)
and tighten securely with a wrench.
Make sure the o-ring on the transducer is in place
i
before threading the transducer into the filter
housing.
11. Push the grommet (6) into the mounting plate.
12. Connect the transducer wire to the motor controller (refer
to the Connection Diagram, section 9.7).
13. Place the motor cover (4) back over the motor controller
(5). Secure with the two motor cover screws (3).
14. Slide the motor shroud (2) over the motor assembly.
15. Secure the motor shroud (2) with the two motor shroud
screws (1).
16. Reattach the filter assembly to the cart.
3
4
1
7
8
MAINTENANCE
2
5
6
27