Descargar Imprimir esta página

HT Instruments EQUITEST Manual De Instrucciones página 29

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
5. Connect the alligator clips of the C7000 cable set to the ends
of the object under test (see Fig. 3)
6. Press the GO/STOP key on the MASTER instrument to activate
the test in Manual mode. The instrument provides the test
current >10A AC and the measurement result is shown on the
display of the MASTER instrument (see related user manual)
7. By selecting the Automatic mode on the MASTER instrument
(measurement activated without using the GO/STOP key) check
that the STATUS LED should be steady orange and connect the
alligator clips to the ends of the object under test (see Fig. 3) to
automatically activate the test
8. To save and manage the result, refer to the user manual of the
MASTER instrument
29
EQUITEST

Publicidad

loading