Pocket Digital Multimeter / English
Function
Range
Resolution
HZ(Current
10kHz
0.001Hz
Frequency)
10- 100.00kHz 0.001Hz
HZ(Voltage
Frequency)
Current sensitivity: >15A ; Voltage sensitivity :>30V
9.999Hz
0.001Hz
99.99Hz
0.01Hz
999.9Hz
0.1Hz
Frequency
9.999kHz
0.001kHz
(electronic)
99.99kHz
0.01kHz
999.9kHz
0.1kHz
9.999MHz
0.001MHz
General Specifications
Clamp jaw opening 0.24" (6mm) approx.
Display (6000 counts) LCD
Analog Bargraph/segments */41 seg.
Continuity check Buzzer sounds at less than 50Ω
Test current of 0.5mA typical; Open circuit voltage
Diode test
≤3VDC typical
Low Battery indication '
' is displayed
Over-range indication ' OL' display
Measurement rate 3 readings per second, nominal
Input Impedance 10MΩ (VDC and VAC)
Operating Temperature 41
o
o
F to 104
F (5
o
o
Storage Temperature -4
F to 140
F (-20
Max 80% up to 87
Operating Humidity
o
o
at 104
F (40
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Battery 2*1.5VAAA Battery
Auto power OFF After approx. 15 minutes
8
Accuracy
±0.1% of rdg ± 3digits
±0.1% of rdg ± 3digits
±(1.0% reading + 5 digits)
o
o
C to 40
C)
o
o
C to 60
C)
o
o
F (31
C) decreasing linearly to 50%
C)
IEC/EN61010-1, IEC/EN61010-2-031, IEC/EN61010-2-032
Safety
Measurement category: CATIII 300V, CATII 600V,
Pollution degree: 2
6. Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove
the test leads from the input terminals, and turn OFF the meter before opening the
case. Do not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives
or so lvents. If the meter is not to be used for 60 days or more, remove the battery and
store it separately.
Battery Replacement
1. Remove the Phillips head screw that secures the rear battery door
2. Open the battery compartment.
3. Replace the 1.5V*2 AAA battery.
4. Secure the battery compartment.
Pocket Digital Multimeter / English
9