Features and Functions

Making Measurements

Auto Identification (Ai) function

Press
device-under-test (DUT).
Figure 2-1
The
– selects an appropriate measurement in the primary display (L, C, or R) and
– selects an appropriate range, and
– selects an appropriate measuring mode (series or parallel).
The Ai function helps to identify L, C, and R measurements automatically
NOTE
according to the angle of impedance detected in the DUT. See
phase angle rules.
The default phase angle condition is set to 10°. You can change this angle in the
Setup menu from 5° to 45°. See
condition"
42
to automatically identify the appropriate measurement required for the
Using the Ai function
annunciator will flash while the LCR meter identifies the DUT, and
secondary display (D, Q, or θ),
on page 77 for more information.
"Changing the Ai function's phase angle
Keysight U1731C/U1732C/U1733C User's Guide