During its entire life cycle, an aerosol head should not be back-
washed more than 25 times with the cleaning aid.
The cleaning aid does not extend the life cycle of the aerosol head.
2 Product description
Intended purpose
The easycare cleaning aid is intended to clean the membrane
of the aerosol head of eFlow
backwashing.
Intended use
The cleaning aid must only be used in accordance with its intended
purpose. No areas of use have been defined for the cleaning
aid, and no contraindications to its use are known. Backwash-
ing unblocks the pores in the membrane of the aerosol head by
flushing with liquid in the opposite direction to the direction for
nebulising.
For a backwashing procedure, you need 5 ml backwash liquid.
Only use one of the liquids listed below as the backwash liquid:
- 5 ml isotonic sodium chloride solution (0.9%)
- 5 ml distilled water
- 5 ml deionized water
Info
Backwashing is no substitute for the steps described in the
eFlow
rapid instructions for use regarding cleaning and
®
disinfection.
10
Technology nebuliser handsets by
®
– 2022-11