Activating A Mechanical Emergency Release; Charging The Battery; Using A Power Cord To Charge The Battery - Invacare Roze Manual Del Usario

Ocultar thumbs Ver también para Roze:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

SECTION 3—OPERATION
Activating a Mechanical Emergency
Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down or lift it up if the hand 
control is not functioning properly.
To activate the primary emergency release, insert a pen into the 
hole labeled Emergency Up or Emergency Down on the control 
box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only be 
used if the primary emergency release procedure is not functioning or is 
unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 
function.
To activate the secondary emergency release, pull up on the RED 
emergency grip and pull down on the boom at the same time.

Charging the Battery

NOTE: There are two different methods to charge the battery. One 
method uses a power cord that attaches to the control box, the other 
requires the battery to be mounted to the battery charger. Follow the 
appropriate procedure to charge the battery for your patient lift.

Using a Power Cord to Charge the Battery

NOTE: For this procedure, refer to FIGURE 3.2.
CAUTION
The emergency stop MUST NOT be activated -
otherwise it will be impossible to charge the
battery. While charging takes place, the patient
lift cannot be used. DO NOT attempt to move
the patient lift without unplugging the power
cord from the wall outlet. DO NOT attempt to
use the patient lift if the battery housing is
damaged. Replace a damaged battery housing
before any further use.
1.
Attach the power cord to the control box.
2.
Plug the power cord into a socket outlet.
NOTE: The battery will charge in approximately 4 hours. Charging 
must be done in a room with good air ventilation.
3.
Disconnect the power cord from the socket outlet after the 
battery has been fully charged.
Roze™Stand Assist Patient Lift
Control Box
Emergency
Up Hole
FIGURE 3.2 Control Box Details
FIGURE 3.3 Secondary Emergency Release
14
Emergency Down
Hole
Maintenance Indicator
(refer to page 23)
Charger Cable
connects here
RED Emergency
Grip
Part No. 1164940