12. Use of the microscope in fluorescence (reflected light)
12.1 Turning on fluorescence LED
Move the main switch in the position "I" to turn on the microscope,
then rotate the adjustment knob ① (Fig. 31) to adjust the
fluorescence light intensity.
12.1.1 Switching fluorescence filter cubes
Move the filter lever (placed on the left side of the microscope)
②
to insert the desired filter cube (see tables below). (Fig. 32)
12.1.2 Available fluorescence filter cubes
•
IM-300LD2
F I L T E R
EXCITATION
NAME
FILTER
B
455-495 nm
G
510-550 nm
•
IM-300LD4 / IM-300LD4D
F I L T E R
EXCITATION
NAME
FILTER
M-1233
325-375 nm
M-1233.1
340-390 nm
M-1232
390-420 nm
M-1230
455-495 nm
M-1230.1
455-495 nm
M-1231
510-550 nm
M-1231.1
510-550 nm
M-1238
582-603 nm
M-1234
590-650 nm
M-1235
595-645 nm
M-1236
623-678 nm
M-1237
720-760 nm
DICHROIC
EMISSION
MIRROR
FILTER
500 nm
510LP nm
570 nm
575LP nm
DICHROIC
EMISSION
MIRROR
FILTER
415 nm
435LP nm
405 nm
420-470 nm
440 nm
450LP nm
500 nm
510LP nm
500 nm
518-542 nm
570 nm
575LP nm
570 nm
585-625 nm
610 nm
615-645 nm
660 nm
665LP nm
655 nm
665-715 nm
685 nm
690-750 nm
770 nm
780LP nm
Page 25
①
②
APPLICATIONS
•
GFP, Alexa Fluor 488, Calcein, SYBR Green,
FITC, Fluo-4, MitoTracker Green
•
Spectrum Gold, Propidium Iodide, Ethidium Ho-
modimer I
APPLICATIONS
•
DAPI, Hoechst, NucBlue, Alexa Fluor 350, BFP,
Coumarin
•
Pacific Blue, Spectrum Blue
•
GFP, Alexa Fluor 488, Calcein, SYBR Green,
FITC, Fluo-4, MitoTracker Green
•
Spectrum Gold, Propidium Iodide, Ethidium Ho-
modimer I
•
Texas Red, Alexa Fluor 594, mRaspberry, Zom-
bie Red
•
Cy5, Alexa Fluor 647, DRAQ5
•
Alexa Fluor 660, DRAQ5
•
Indotricarbocyanine, DiR
F ig. 31
F
ig. 31
F ig. 32
F
ig. 32