Table 2: Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the devices should assure that it is used in such an environment.
IEC 60601
Immunity Test
Test Level
± 8 kV contact
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 15 kV air
Electrical fast
± 2 kV for power
transient/burst IEC
supply lines
61000-4-4
± 1 kV for
input/output lines
Surge
2 kV
IEC 61000-4-5
0, 90, 180 & 270
degree phase angles
Voltage dips,
100% drop, 0.5 periods
short
(0°, 45°, 90°, 135°,
interruptions and
180°, 225°, 270°, 315°)
voltage variations
on power supply
input lines
100% dip, 1 period
IEC 61000-4-11
30% dip, 25/30
periods
100% drop, 5 seconds
3 A/m
Power
frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
Note
UT is the a.c. mains voltage prior to application of the test level.
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
NA
2 kV
0, 90, 180 & 270
degree phase angles
100% drop, 0.5 periods
(0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°)
100% dip, 1 period
30% dip, 25/30
periods
100% drop, 5 seconds
3 A/m
25
Electromagnetic
Environment
Guidance
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the device
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.