Warning/Fault
Warning LED
+ audible warning
Warning LED
+ audible warning
1630732-A
Cause
Pressure too low
Alternating Mode Failure
(no alternation)
Solution
1. Reset the audible warning – turn off Power and press the "Audible
Warning Mute" button.
2. Check that the handle is intact, ensuring all four sealing connectors
are firmly fitted to Control Unit and the air hoses. Check the CPR
Tag is attached and all four sealing connectors are firmly secure.
3. Check all air hoses along the inside of the mattress – each should
be firmly connected. Check each air cell is securely attached to
its connecting air pipe.
4. Check all cells, pipes and hoses for any air leakage.
5. Check that the air filter cover is correctly secured and the air
filter is clean.
6. Switch on Power.
1. Reset the warning – turn off Power and press the "Audible
Warning Mute" button.
2. Disconnect the air hoses to reduce pressure – reconnect when
pressure has decreased.
3. Restart the system by switching the power button off and on.
The system automatically stats in Static Mode and will switch to
Alternating mode when operational pressure is reached.
4. Reset the personal settings.
Troubleshooting