A Power button
Turns system power on and off. Press for three (3)
seconds
B Control unit Lock / Unlock button
Press for at least three (3) seconds to lock the control
unit settings – a beep sounds and the amber LED
illuminates to indicate system is locked. When locked,
all the buttons will be locked except for the lock
button. Press again for at least three (3) seconds to
unlock (beep sounds and amber LED turns off).
C Audible Warning Mute button
Silences the audible warning (on / off). Audible
warning will resume after 17±3 minutes if cause of
failure is not resolved.
1630732-A
J
E
D
C
F G H I
D Max Firm button
Press to facilitate rapid inflation to maximum pressure
setting (60 mmHg). To disengage maximum pressure
press again for 3 seconds. If not disengaged the
system will automatically revert back to alternation
mode with medium pressure after 30±3 minutes, for
patient safety.
E Mode button
Press to select either Alternation Mode (alternative
cells cyclically inflating and deflating) or Static Mode
(all cells fully inflated with no dynamic alternation).
In order to activate or inactivate CLP function, press
the button for 2 seconds.
F Alternation failure indicator
See section "Audible Warning functions" for more
information.
G Low pressure indicator
See section "Audible Warning functions" for more
information.
H Service indicator
See section "Audible Warning functions" for more
information.
I Power cord indicator
See section "Audible Warning functions" for more
information.
Usage
17