Display
Audible Warning
Signal
Alternating Mode
failure
Low Pressure
1630732-A
Description
Indicates with an audible
sound (press mute button
to silence) and a flashing
LED light when there is no
alternation. The LED light
will change to a fixed light
within 5 min.
Indicates that pressure has
fallen below the tolerance
limit (below setting
pressure –5 mmHg) or
more. This will trigger the
audible warning (press
mute button to silence)
and a flashing LED light.
The LED light will change
to a fixed light within 7
min.
Display
Audible Warning
Signal
Power failure
Initial failure (the
mute button LED
will be constantly
lit)
Service indicator
If Audible warning activates and the system fails to
inflate or loses pressure, refer to section "Identifying
and repairing faults".
Usage
Description
Indicates no pressure
output due to mains
power failure. This
will trigger the audible
warning (press mute
button to silence) and a
flashing LED light. The LED
light will change to a fixed
light within 3 seconds.
When the power is back
on, the audible warning
and the LED light will
revert to normal mode.
Indicates that the system
did not complete
initialization within 50
Min, the mute button LED
light will have a fixed light.
Indicates that service is
needed, the LED light will
have a fixed light. This will
occur when the control
unit has been running
for approximately 9.000
hours.
21