IB 3BU1-5(new address)
2003.8.21
6.1. How to print the last stored measurement
(Single data print)
a) Select correct printer objective first before you press
the PRINTER button.
b)Connect the printer with the monitor and
switch the printer on (for details please
refer to the manual of the printer).
c) Press the PRINTER button of the monitor once. The display will show
"Pr1" and the printer will print the last stored measurement for User 1
or 2 based on your selection.
6.2.How to print all measurements stored in memory and a data
chart, (data print out with chart).
a) Select correct printer objective first before you depress the PRINTER
button.
1
b) Connect the printer with the monitor and switch the printer on (for
details please refer to the manual of the printer).
18
8:42 AM
Page 18
c)Depress the PRINTER button for at least
more than 3 seconds. The display will
show "Pr1" followed by "Pr2"
d) Release the button, and the printer will print all stored measurements
with a corresponding chart.
6.3. Stop printing
If you would like to stop printing, click the
PRINTER button during the printin
process.
NOTE:
Further details regarding the printer can
be found in the printer instruction
manual.
1
7. Error messages/malfunctions
If an error occurs during a measurement,
the measurement is discontinued and a
corresponding error code is displayed
(Example: Error No. 2).
Error No.
Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason: The arm
2
was moved during the Measurement (Artefact).
ERR 3
The inflation of the cuff takes too long. The cuff is not correctly seated.
ERR 5
The measured readings indicated an unacceptable difference between systolic and
diastolic pressures. Take another reading following directions carefully. Contact you
doctor if you continue to get unusual readings.
1
2
1
19