Correct maintenance of the autoclave assures correct funcioning of it and a secure saving
in terms of time and costs for assistance and maintenance. The following operations are
compulsory feasible by operators.
Cleaning of chamber
Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the
introduction in the discharging circuit of material which can cause obstructions. For correct
cleaning use only deminerlized water and the abrasive sponge provided (non abrasive side
– Fig. 5).
To execute absolutely with cold chamber to avoid burns – Never use solvents, detergents,
chemical solutions, descaling agents or other similar products.
Cleaning of chamber filter
Pull up the filter (spare code DXBA091), paying attention not to damage it, wash with
demineralized water and dry with cloth. Then replace in the seat, making sure that
protrudes about 15 mm.
Cleaning of tray and tray holder
Clean with sponge supplied (non abrasive side) and demineralized water.
Change of the bacteriological filter
Substiture the bacteriological filter (Fig.A–pos.
and clockwise to screw. Use only original filter (spare code DAVA101). It's necessary to
remember to reset the cycles counter when you change the filter selecting RESET
COUNTER in the BACTERIOLOGIC FILTER submenu.
Cleaning of the door gasket
Periodically remove eventual residuals which are deposited on the circumference of the
seal (spare code DANA038) using water and the sponge provided (non abrasive side), or
else a moistened cloth.
36
MAINTENANCE
Every 200 cycles or when it assumes a dark colour
13
) turning it counterclockwise to unscrew
Every 20 cycles or one a week
Every 20 cycles or once a week
Every 20 cycles or once a week
Every 20 cycles or once a week