ENGLISH
14.1 Authorized technicians ordinary maintenance
Clean the chamber
Clean the metal filter in the chamber
Replace the seal of the chamber door (DANA038)
Lubricate the closing system of the chamber door
Adjust the closing system of the chamber door
Replace the bacteriological filter (DAVA101)
Replace the water loading filter (DARA054)
Clean the Y brass filter downstream of the radiator
Clean the water loading solenoid valve or replace it if necessary
Clean the tanks
EVERY ONE
Check the tank level probes
YEAR /
Clean the radiator and cooling fans
EVERY 800
Check the condition of the electrical/pneumatic circuit
CYCLES
Check that there are no traces of disinfectant in the pneumatic circuit.
Please note: ignoring traces of disinfectant may result in the need for
supplementary maintenance of the equipment
Check the safety valve
Check the efficiency of the vacuum pump
Perform validation in compliance with the local standards and on the basis
defined by the local standards
Perform the electrical safety tests in compliance with the local standards
and on the basis defined by the local standards
Measure the conductivity of water
Replace the valves and o-rings of the vacuum pump – Clean/check the
condition of the Knf membranes (CPGM025 x4 – CPGM043 x2) -
AFTER 1600
Thomas (CPGM050 x4 – CPGM051 x4 – CPGM052 x2)
CYCLES
Check the three N/C solenoid valves
Check the heating part
Replace the three N/C solenoid valves (CEECG021 x3)
AFTER 2400
Replace the heating part (DXBA835 or DHYA035)
CYCLES
Clean/check the condition of the membranes of the vacuum pump
38
......................
S/cm