Increasing and Decreasing the Amplitude

The amplitude of stimulation is the peak value of a stimulation pulse and is measured in
milliamps (mA).
To increase or decrease the amplitude of the selected area:
1
To increase the amplitude, click or click and hold one of the up arrows above the AMP label
on the Programming screen.
2
To decrease the amplitude, click or click and hold one of the down arrows above the
AMP label..
Click...
To increase or decrease the amplitude for all areas that have stimulation turned on:
1
On the Programming screen, click the
If the Global button is selected (the button will have a bright blue circle around it), only the
amplitude can be adjusted.
2
Increase or decrease the amplitude as described above.
3
Click the
Vercise™ DBS Programming Manual
91053785-07
26 of 323
to...
Increase the amplitude quickly.
Increase the amplitude slowly.
Decrease the amplitude slowly.
Decrease the amplitude quickly.
button again to discontinue Global amplitude adjustment.
button.