Usage
If necessary, correct the X-ray positioning
i
beam manually�
EN
US
Once the patient has been correctly posi-
i
tioned using the positioning beams, deacti-
vate the positioning beams using the button�
Adjusting the head supports
Adjust the height of the head supports�
i
Carefully press the head supports by hand to-
i
wards the head in order to check if they are in
the right position� The device or the head sup-
ports are not damaged in the process�
Ideally, the head supports should make con-
tact slightly above the eye brows; correct the
position as required�
32
OBJ_DOKU-0000003575-mehrere-005.indd 32
Close the head supports with the
i
To do this, just press the button briefly – don't
press and hold�
The head supports automatically touch
against the head of the patient at a defined
pressure�
7.5 Start a test run
The test run ensures that the unit can perform
the image acquisition without any problems�
This prevents unnecessary exposure of the pa-
tient to radiation�
No radiation is generated during the test
run�
Requirements:
– The patient is correctly positioned in the unit
using the positioning aids and the positioning
beams�
– The required imaging program has been
selected�
Touch Test run on the touch screen�
i
Tap Run and hold�
i
While doing this, constantly monitor the move-
ments of the unit� If the unit is obstructed in its
movements, release the Run button� The unit
stops immediately� Reposition the patient�
Press Return run to perform the return run�
i
2210200496L41 1904V004
button�