7.6 Recording the X-ray image

CAUTION
Injuries due to X-rays
X-rays can cause tissue damage�
Comply with the radiation protection
i
regulations�
Maintain the minimum distance�
i
CAUTION
Danger of excessively high radiation
dose
Before an image acquisition is trig-
i
gered, all data entered on the PC
must be checked on the touch
screen�
Check all parameters on the touch screen and
i
change them if necessary�
The changed parameters are immediately
synchronized with the imaging software� The
parameters can then no longer be changed in
the imaging software�
Remind the patient to press his/her tongue
i
against the roof of the mouth during image
acquisition�
Press Start to confirm the parameters�
i
The C-shaped arm is being positioned� The
LED on the exposure switch and the status
LED on the unit light up green�
The touch screen displays that the unit is
ready to take an image�
2210200496L41 1904V004
OBJ_DOKU-0000003575-mehrere-005.indd 33
Trigger the image by pressing and holding the
i
button on the exposure switch until the
acoustic signal stops and the control lamp
goes out� The scan times depend on the pa-
tient type, imaging program and image quality
(see "12 Program parameters")�
Image acquisition is started� During the ac-
quisition, the LED on the exposure switch
and on the unit illuminates yellow� An acous-
tic signal is issued�
If the button on the exposure switch
is released before the control lamp
goes out or the emergency stop
switch is pressed (e�g� if there is a
danger to the patient or to anyone
else in the area), the ongoing image
acquisition is aborted� The X-ray im-
age will be unusable as a result and
should be retaken as required� In this
case, the operators must use their
skills and training to decide on the
risks of a repeated image acquisition�
In addition, an error messages ap-
pears on the touch screen�
While an X-ray is being taken, this is indicat-
ed on the touch screen by:
After the acquisition of maxillary joint
images, it is necessary to confirm a
message on the touch screen and
trigger a second image acquisition�
The images are then combined into a
single image�
The LED on the unit lights up blue when the
X-ray acquisition has been completed�
The C-shaped arm does not automatically
move back to the starting position after the
trigger button is released�
Usage
EN
US
33
17.04.2019 14:29:02