Figure 3. Tighten the setscrew clockwise
Remove the torque wrench and check the septum to ensure that it closed. If the septum did not
4.
close, gently reseat the septum flaps.
If implanting two leads, repeat the previous steps. If implanting a single lead only, insert the
5.
header port plug into the unused port, and use the torque wrench to tighten the setscrew until it
clicks.
Figure 4. Insert the port plug
Implanting the IPG
The following steps outline the suggested procedure to implant the IPG:
Place the IPG into the IPG pocket with the logo side facing the skin surface and at a depth not
1.
to exceed 4.0 cm (1.57 in).
NOTE: By implanting the IPG with the logo side facing the skin surface, you enhance the
IPG's ability to detect a magnet.