The IPG has the following operating parameters.
Table 4. Operating parameters for the IPG
Parameter
Tonic Range
Pulse width
20–1000 µs
Frequency
2–200 Hz
200–500 Hz
500–1200 Hz
Burst rate
—
frequency
Intraburst
—
frequency
Amplitude
0–25.5 mA
NOTE: Columns with * represent operating parameters for burst programs on IPGs capable of burst
stimulation mode.
NOTE: The number of stim sets in use for a tonic program governs the maximum frequency
(1200/number of stim sets).
NOTE: The maximum current depends on the impedance, frequency, and pulse width settings.
Compatibility Guidelines for IPGs with Compatible Headers
IPGs with compatible headers are compatible with the following Medtronic™ leads and extensions
available before May 15, 2015. (Medtronic is a trademark of Medtronic, Inc.)
Table 5. Compatible Medtronic leads and extensions
Device
Model
Permanent lead
3776-45, 3776-60, 3776-75, 3876-45, 3876-60, 3876-75, 3777-45, 3777-60,
3777-75, 3877-45, 3877-60, 3877-75, 3778-45, 3778-60, 3778-75, 3878-45,
3878-60, 3878-75, 39286-30, 39286-65, 39565-30, 39565-65
Extension
3708120, 3708140, 3708160, 3708220, 3708240, 3708260, 3708320,
3708340, 3708360
Appendix B: System Components and Accessories
The Proclaim™ neurostimulation system includes the following components.
NOTE: Not all models are available in all countries. Contact your local representative for more
information.
ARTMT100140147B.pdf 16
Tonic Steps
Burst Range*
10 µs
20–1000 µs
(20–500 µs range)
50 µs
(500–1000 µs
range)
2 Hz
—
10 Hz
—
20 Hz
—
—
10–60 Hz
—
250–500 Hz
500–1000 Hz
0.1–1.0 mA
0–12.75 mA
12
Burst Steps*
50 µs
—
—
—
10 Hz
10 Hz
20 Hz
0.05–1.0 mA
5/2/2016 3:20:10 PM