1.3 Conditions of Use

INFORMATION
Advise your patients of the information in this section.
The C-Leg® was developed for everyday activities and not for extreme sports such as free climbing, parachuting or
paragliding. For the necessary environmental conditions, please see Section 3.3 "Technical Information".
The C-Leg® prosthesis system is designed exclusively for use on one patient. Use of the product by other persons
is not approved by the manufacturer.

1.4 Qualification of the Prosthetist

The fitting of a patient with a C-Leg® may only be carried out by prosthetists who have been authorized with the
corresponding Otto Bock training.

1.5 Safety Instructions

INFORMATION
Advise your patients of the information in this section.
CAUTION
Non-observance of safety instructions. Failure to follow the below-mentioned safety instructions can lead to
a faulty control or malfunction of the C-Leg® and result in risk of injuries for the patient as well as damage to the
C-Leg®.

General Safety Instructions

CAUTION
Alignment and adjustment error. During alignment and adjustment of the prosthesis, errors can occur result-
ing in malfunction of the joint up to loss of function due to structural failure. This can cause the patient to fall.
• Participation in an Otto Bock product training course for the C-Leg® is obligatory before the first fitting.
Additional product training courses may become necessary to qualify for fitting product updates.
• The tubes must not be shortened as that may damage the integrated strain gauges.
• During the data transfer (PC to C-Leg®), the prosthesis wearer must remain standing or sitting still and the
BionicLink or the communication cable must not be removed.
• If the prosthesis wearer uses crutches or walking canes during adjustment, readjustment is required as soon
as he stops using these walking aids.
CAUTION
Use of inappropriate prosthetic components. If inappropriate prosthetic components are installed in the
prosthesis, malfunction of the joints can occur up to loss of function due to structural failure. This can cause the
patient to fall.
Combine the C-Leg® only with adapters and feet approved by Otto Bock (see component overview on the folded
front page).
31