2.6 BionicLink/Optional: Serial Data Cable
INFORMATION
Prior to connecting the 60X3 BionicLink, read the corresponding 647G192 Instructions for Use.
1
1 Remove the protective plug (4X73) and connect the wireless adapter cable (BionicLink) (optional: connect serial
data cable).
2 Snap the BionicLink into place and attach it medially (for left-side fittings) or laterally (for right-side fittings) to the
C-Leg®.
2.7 C-Soft
INFORMATION
Prior to the parameter adjustment with 4X180=* C-Soft, read the corresponding 647G268 Instructions for Use.
INFORMATION
Otto Bock Data Station is the platform for Otto Bock applications such as C-Soft and others.
Otto Bock C-Soft cannot be used without the Otto Bock Data Station. The Otto Bock Data Station is a standard
component of C-Soft and is automatically installed when installing C-Soft.
2.8 Optional: Parameter Settings with Foam Cover and C-Soft
The 3S26 or 3R59 Foam Covers can be used for the cosmetic enhancement of the leg prosthesis.
For that purpose, a 4X78 Charger Extension Cable can be integrated into the C-Leg® fitting. If the charger exten-
sion cable is not used, a connection cap made of 2 mm Pedilin® makes it possible to pull down the foam cover in
order to reach the charging receptacle at the C-Leg®.
Allow a 60 mm "compression allowance" to minimize the effect of the protective foam cover on the knee function.
When determining the proper cover length, increase the thigh dimension by 30 mm and the shin dimension by the
same amount.
The mounted foam cover changes the damping behaviour. With the 60X3 BionicLink (wireless programming inter-
face), which is attached below the pre-shaped foam block, the damping behaviour can be completely taken into
account by changing the parameter settings.
That process also requires the 60X5 (60X4) BionicLink PC for the data transfer and the 4X180=* C-Soft for the
parameter adjustment (see Section "2.7 C-Soft" ).
42
2
3
4