5 General Technical Specifications

AC40 Technical Specification
Safety Standards
EMC Standard
Audiometer Standards
Calibration
Air Conduction
Bone Conduction
Free Field
High Frequency
Effective masking
Transducers
Patient Response switch
Patient communication
Monitor
Special tests/test battery
(some are optional)
Stimuli
Tone
Warble Tone
IEC60601-1:2005; ES60601-1:2005/A2:2010; CAN/CSA-C22.2 No. 60601-1:2008;
IEC60601-1:1988+A1+A2
Class I
Type B Applied parts
IPx0
IEC 60601-1-2:2007
Tone: IEC 60645-1:2012/ANSI S3.6:2010 Type 1-
Speech: IEC 60645-2:1993/ANSI S3.6:2010 Type A or A-E
Calibration information and instructions is located in the AC40 Service manual
TDH39:
DD45:
HDA300:
HDA280
E.A.R Tone 3A/5A:
CIR 33
IP30
B71:
B81
Placement:
ISO 389-7 2005, ANSI S3.6-2010
ISO 389-5 2006, ANSI S3.6-2010
ISO 389-4 1994, ANSI S3.6-2010
TDH39
DD45
HDA300
HDA280
DD450
B71 Bone
B81 Bone
E.A.R Tone 3A/5A
CIR 33
IP30
Two push button.
Talk Forward (TF) and Talk Back (TB).
Real stereo output through built-in speakers or through external earphone or assistant
monitor.
Stenger
ABLB
Weber
Tone decay
Langenbeck (tone in noise).
Masking Level Difference
Pediatric Noise Stimuli
Multi Frequency
High Frequency
Speech from Hard-drive (Wave Files)
SISI
Master Hearing Aid
Hearing Loss Simulator
QuickSIN(tm)
Auto threshold:
Hughson Westlake
o
Békésy
o
125-20000Hz separated in two ranges 125-8000Hz and 8000-20000Hz.
Resolution 1/2-1/24 octave.
1-10 Hz sine +/- 5% modulation
ISO 389-1 1998, ANSI S3.6-2010
PTB/DTU report 2009
PTB report PTB 1.61 – 4064893/13
PTB report 2004
ISO 389-2 1994, ANSI S3.6-2010
ISO 389-2 1994
ISO 389-2 1994, ANSI S3.6-2010 DES-2361
ISO 389-3 1994, ANSI S3.6-2010
ISO 389-3 1994, ANSI S3.6-2010
Mastoid
Headband Static Force 4.5N ±0.5N
Headband Static Force 4.5N ±0.5N
Headband Static Force 8.85N ±0.5N
Headband Static Force 5N ±0.5N
Headband Static Force 10N ±0.5N
Headband Static Force 5.4N ±0.5N
Headband Static Forces 5.4N±0.5N