The speech screen set up in table mode using wave files (26) - under Setup (19).
Function Key
10
11
12
13
14
15
16
17
When the Wavefile test is started, the F-buttons will change to recording mode.
In recording mode, if the protocol has been set to continue/timeout after the word has been played the
word will be grey colored, waiting for the operator's input.
The input can either be Corret(56) / Incorrect(55) on the keyboard or by using the Phoneme score on the
F-buttons. The tested can be paused on the play/pause button.
If the recording mode has been set to manual, the words can be selected, one by one, by using the
forward/reverse button on the F-buttons, press play to play the word.
Description
Only available if High Frequency is available (optional license) on the
instrument. Selects the HF phone connected to the separate HF connectors.
Choose between HL, MCL and UCL by holding the Function Key (10) and
select the required measuring type by using one of the rotary wheels (56)/(57).
The condition under which the speech test is done: None, Aided, Binaural or
Aided & Binaural.
Switch between magnified top bar and normal sized top bar.
Use the HL dB dial (57) to select the different items from the lists:
Different lists can be changed in the "List" option. Use the HL dB dial (57) to
select the different items from the lists.
Start playing the wave files.
Stop playing the wave files.