Green: Charging complete
Yellow: Failure Mode.
If the red LED does not light and remain ON when the charger is
attached, it is possible:
an incomplete electrical connection exists between the charger
and the Lithium Ion battery pack contact points OR
the temperature of the battery pack is outside of the previously-
indicated range.
During recharge, the red LED turning OFF and and green LED not
lighting is an indication of an improper charging process.
This is most likely caused by a battery pack temperature being
outside of the previously-indicated range.
Restart the charging process with the battery pack in a different
temperature environment.
Failure mode, indicated by the yellow LED, will occur if either:
the battery pack has been deeply depleted beyond the point
where it will accept a charge
an internal charger fault has been detected, which would result
in an erroneous charging condition.
Charger may be left on the instrument when charge is complete.

Changing Instrument Settings

Many options can be set using the instrument buttons.
If the Sirius Multigas Detector was ordered with the optional
datalogging, the MSA FiveStar LINK software can be used to set
most of the instrument selections, including some that cannot be
changed from the instrument's front panel buttons.

Accessing the Instrument Setup Mode

FIGURE 4-3 shows how to enter the Set-up mode.
4-4