Table 4 Block 3 Analyzer ERROR
Bit Error
Note
0
Measurement Calibration
An error occurred during the last calibration.
Error
1
Electronic Adjustment
An error occurred during the last electronic
Error
calibration.
2
Cleaning Error
The last cleaning cycle failed.
3
Measuring Module Error
A failure has been detected in the Measuring
Module.
4
System Re-initialization
Some settings have been detected as
Error
inconsistent and have been set to the factory
defaults.
5
Hardware Error
Any hardware error has been detected.
6
Internal Communication
A communication failure within the device has
Error
been detected.
7
Humidity Error
Excessive humidity has been detected in this
device.
8
Temperature Error
The temperature within the device is more
than the specified limit.
9
Fixed at zero.
10
Sample Warning
Some action is necessary within the sample
system.
11
Questionable Calibration
The last calibration may not be accurate.
Warning
12
Questionable
One or more of the measurements of the
Measurement Warning
device may not be accurate (Quality bad or
out of range).
13
Safety Warning
A condition has been detected that may result
in a safety hazard.
Table 4 Block 3 Analyzer ERROR (continued)
Bit Error
Note
14
Reagent Warning
Some action is necessary with the reagent
system.
15
Maintenance Required
Maintenance is necessary on this device.
Warning
Table 5 Block 3 Analyzer STATUS 1
Bit Status1
Note
0
Calibration in progress
The device has been put in a calibration
mode. The measurements may not be valid.
1
Cleaning in progress
The device has been put in a cleaning mode.
The measurements may not be valid.
2
Service/Maintenance menu
The device has been put in a service or
maintenance mode. The measurements may
not be valid.
3
Common Error
The device recognizes an error. See the error
registers for the error class.
4
Measurement 0 Quality Bad The precision of the measurement is out of
the specified limits.
5
Measurement 0 Low Limit
The measurement is less than the
measurement range.
6
Measurement 0 High Limit
The measurement is more than the
measurement range.
7
Measurement 1 Quality Bad The precision of the measurement is out of
the specified limits.
8
Measurement 1 Low Limit
The measurement is less than the
measurement range.
9
Measurement 1 High Limit
The measurement is more than the
measurement range.
10
Measurement 2 Quality Bad The precision of the measurement is out of
the specified limits.
English 11