When the manual configuration is selected, the telegram data structure
can be configured by the user (refer to
Table 2 Profibus data telegram structure for digital sensors
Byte number
Data
1–2
Classified error
3–4
Classified status
5–8
Measurement 1
9–12
Measurement 2
13–16
Measurement 3
Table 3 Profibus data telegram structure for analyzers
Byte number
Data
1–2
Analyzer error
3–4
Analyzer status 1
5–6
Analyzer status 2
7–10
Measurement 1
11–14
Measurement 2
15–18
Measurement 3
19–22
Measurement 4
23–26
Measurement 5
27–30
Measurement 6
10 English
Setup the network
on page 8).
Data type
Integer (2 bytes)
Integer (2 bytes)
Floating (4 bytes)
Floating (4 bytes)
Floating (4 bytes)
Data type
Integer (2 bytes)
Integer (2 bytes)
Integer (2 bytes)
Floating (4 bytes)
Floating (4 bytes)
Floating (4 bytes)
Floating (4 bytes)
Floating (4 bytes)
Floating (4 bytes)
Display values
The Profibus data block structure
changes in the PLC configuration.
The primary value is always the measured value.
The secondary value, if not available, is filled with 0xFF.
The tertiary value, if not available, is filled with 0xFF.
Figure 8 Profibus message data block structure

Process data controller block

The data block for the analyzer is similar to the data block for devices.
The structure of the analyzer data block is independent of the number of
connected devices:
• analyzer_ERROR
• analyzer_STATUS 1
• analyzer_STATUS 2
• Primary value
• Secondary value
• Tertiary value
Table
4,
Table 5
and
Table 6
show the data definitions for error, status
1 and status 2 in the analyzer.
(Figure
8) can replace devices without

Publicidad