Remove The Access Plugs; Electrical Installation - Hach LANGE Manual Del Usuario

Ocultar thumbs Ver también para LANGE:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
Figure 2 Product components
1 Profibus network card
2 Washer nut (2x)
Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.

Electrical installation

D A N G E R
Electrocution hazard. Always remove power to the instrument before
making electrical connections.
D A N G E R
Electrocution hazard. Use only fittings that have the specified
environmental enclosure rating. Obey the requirements in the
Specifications section.
3 Ring terminal (2x)
W A R N I N G
Electrical shock hazard. Externally connected equipment must have an
applicable country safety standard assessment.
N O T I C E
Make sure that the equipment is connected to the instrument in accordance with
local, regional and national requirements.

Remove the access plugs

Install cables and conduit through the electrical access ports. Refer to
Figure
3. Remove rubber sealing plugs by pushing them out from inside
the enclosure to unlock the seal, and then remove completely by pulling
from the outside. Remove knockouts as necessary from the electrical
access plate with a hammer and screwdriver. To keep the enclosure
rating, put a cover on all ports that are not used.
Figure 3 Electrical access ports
1 Power in (power cord only), no
ground plate. Do not use for
conduit.
2 Communication and network
modules (3x)
3 Communication and network
modules (8x)
4 Power in or out (conduit or power
cord), ground plate, communication
and network modules (8x)