Event Log; Replacement Parts - Hach LANGE Manual Del Usuario

Ocultar thumbs Ver también para LANGE:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
Table 7 Error messages (continued)
Bit Message
14
Reagent warning
15
Maintenance required
warning
Table 8 Status indicator messages
Bit Message
0
Calibration in progress
1
Cleaning in progress
2
Service/Maintenance menu
3
Common error
4
Measurement 0 Quality Bad Precision of measurement is out of specified
5
Measurement 0 Low Limit
6
Measurement 0 High Limit
7
Measurement 1 Quality Bad Precision of measurement is out of the
8
Measurement 1 Low Limit
9
Measurement 1 High Limit
10
Measurement 2 Quality Bad Precision of measurement is out of the
11
Measurement 2 Low Limit
12
Measurement 2 High Limit
13
Measurement 3 Quality Bad Precision of measurement is out of the
14 English
Indication
The reagent system requires attention
The device requires maintenance
Indication
The device is in a calibration mode.
Measurements may not be valid.
The device is in a cleaning mode.
Measurements may not be valid.
The device is in a service or maintenance
mode. Measurements may not be valid.
The device has recognized an error. See
Error Register for Error Class.
limits.
Measurement is below the specified range.
Measurement is above the specified range.
specified limits.
Measurement is below the specified range.
Measurement is above the specified range.
specified limits.
Measurement is below the specified range.
Measurement is above the specified range.
specified limits.
Table 8 Status indicator messages (continued)
Bit Message
14
Measurement 3 Low Limit
15
Measurement 3 High Limit

Event Log

Refer to
Table 9
for diagnostic device information.
Table 9 Event log
Event
Description
ADDRESS
Adjusted Profibus address
DATA ORDER
Indicates the data order of 2 word variables in the cyclic
and acyclic Profibus telegram
SIMULATION
Indicates if the simulated data is set into the cyclic
Profibus telegram.
SENSOR POWER
Timestamp of the Profibus card power up
SET DATE/TIME
Setup timestamp of the internal timer of the Profibus card
NEW CONFIG
Timestamp of a new configuration
AUTO CONFIGURE Timestamp of a new menu setting
CODE VERSION
Timestamp of a new software download (Software
version)

Replacement parts

W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment malfunction.
The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Indication
Measurement is below the specified range.
Measurement is above the specified range.