Battery charging

The capacity of the battery is not negatively affected if it is charged when partially empty and do
not need to be fully charged at each occasion.
The battery can be charged when appropriate, or in accordance with alarm signal for
"Battery low" or "Battery empty" and follow instructions in the pump display.
There is approximately 30 minutes of charge remaining on the battery at the onset of the
"Battery Low" warning.
If the battery is completely empty it will take 3 to 4 hours to charge to full status.
If the Invia Liberty Pump is fully charged AND the pump is still connected to an electrical source,
a lightning bolt will appear in the battery icon.
If the pump is disconnected from an electrical source, 4 bars are visible in the battery icon which
indicates that the battery is fully charged.
If Invia Liberty Pump has not been in use, the battery must be charged approximately once every
6 months to ensure optimum function.

To disconnect Invia Liberty Pump from the electrical source

Remove the charger connector and close the safety cover. Invia Liberty Pump switches off
automatically.
WARNING
Do not pull on the cable or the anti-bend protection.
28
To charge the battery place the Invia Liberty Pump in the
Docking Station or plug in the charger to the electrical
outlet port on the pump. You can continue to use the
pump while it is charging.