15 Battery data and test
(AC / DC versions only)
CAUTIONS
Only charge the rechargeable battery when the acoustic signal sounds and the charging
indicator blinks slowly. The rechargeable battery is charged during mains operation.
Duration of pumping
If the battery is fully charged, the operating time is about:
– low vacuum versions: approx. 120 minutes
– high and medium vacuum versions: approx. 30 minutes.
Charging time (charged up to 100 %)
– low vacuum versions: approx. 5 hours
– high and medium vacuum versions: approx. 5 hours.
Testing the rechargeable
battery
1.
Make sure that the rechargeable battery is fully charged.
2.
Pull out the mains plug from the fixed mains socket.
3.
4.
or
5. Measure the time until the pump comes to a stop. The
battery is ok, when the following values are reached:
– low vacuum versions: > 120 min.
– high and medium vacuum versions: > 30 min.
If the values are not reached, repeat test or calibrate
the battery (see next page).
3.1 Press vacuum regulator
and turn it to the right to
select maximum vacuum.
4.1 Switch on the Vario. The
pump will start running.
Caution: c / i versions
must be switched on in
«constant» mode.
Battery data and test
| 25