Installation Procedure; Panel-Mount Stiffener Bracket - Mettler Toledo IND560 Serie Guia De Instalacion

Ocultar thumbs Ver también para IND560 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

IND560xx Terminal Division 2 and Zone 2/22 Installation Manual
Figure 2-4: Grounding Points, Panel Mount (top) and Harsh Enclosure (bottom)

Installation Procedure

Once the information in this chapter and in all other suggested regulatory
documents has been read and understood, the IND560xx terminal may be
installed. Refer to Chapter 2 of the standard IND560 Installation Guide
(#71209395) for details on how to mount the harsh or panel-mount enclosures
and how to make the required terminal block connections for the power, scale and
other accessories. Special installation requirements for the IND560xx terminals are
described below. They are also listed in the Special Conditions for Safe Use section
of this manual.
In addition to the information in this chapter, instructions, control drawings,
installation drawings and details listed in the certificates found in Chapter 3 and
Appendix A of this guide must be followed during the installation.

Panel-Mount Stiffener Bracket

When installing the panel-mount IND560xx terminal, note that the panel stiffener
bracket included with the terminal must be used to guarantee the required IP rating.
After inserting the panel enclosure through the mounting hole in the panel, slide the
stiffener bracket over the housing, then install and tighten the two mounting
clamps. The stiffener bracket is shown installed in Figure 2-5.
2-9

Publicidad

Tabla de contenido
loading

Tabla de contenido