Turn the meter auto-shutoff feature OFF .
Set the date and time on the meter . See the Time and Date
Settings section for details on setting this parameter .
Set the baud rate on the meter to 1200 .
Alternatively, the meter baud rate can be set to 9600 and the
autosampler dip switches can be set to 9600 for the Com 1
port . This is recommended when a printer or computer will
be connected to the autosampler . See the Autosampler Dip
Switch Settings section for instructions .
Set the data output on the meter to printer or computer format .
3 . Review and change the parameters for the autosampler in the setup menu .
Turn the autosampler function On or OFF .
Select the beaker tray that will be used . Either the 28 beaker
or 48 beaker tray can be selected .
Enter the number of rinse beakers . One to five rinse beakers
can be used .
Enter the electrode rinse time in each rinse beaker . The rinse time
can be set from 5 to 60 seconds .
If a pH electrode will be calibrated on the autosampler, enter
the number of pH calibration points . Zero to three points can be
selected . The meter must perform an automatic calibration with
the autosampler, so only pH buffers in the selected buffer set
can be used . Set the number of calibration points to zero if a pH
calibration with the autosampler is not needed . This menu will
only be shown if the meter has pH function .
Thermo Scientific Orion Star
Orion Star Plus Benchtop Meter and Autosampler Interfacing
and Star Plus Meter User Guide
™
EN-77