Ise Temperature Display And Calibration - Thermo Scientific Orion Star Manual Del Usario

Ocultar thumbs Ver también para Orion Star:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

ISE Technique

ISE Temperature Display and Calibration

ISE Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen .
To view the temperature for the ISE measurement line:
1 . Press and hold
2 . Press
to choose the top display line . The arrow icon will point to the
selected line .
3 . Press
/
can be changed to display pH (pH), millivolts (mV), relative millivolts (RmV),
concentration (ISE), temperature (no icon) or a blank line .
ISE Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature
on each measurement line to be manually adjusted .
To calibrate the temperature for the ISE measurement line:
1 . In the measurement mode, press
and press
2 . Press
to begin the calibration .
3 . When the reading stabilizes, the arrow icon and the first digit will flash .
Enter the temperature by pressing
move to the next digit .
4 . Press
to save and end the calibration .
EN-48
until the meter displays the measurement mode .
to change the value on the selected line . The top line
/
until the temperature is shown for the selected line .
Thermo Scientific Orion Star
to choose the top measurement line
/
to adjust each digit and
and Star Plus Meter User Guide
to

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Star plus meter

Tabla de contenido