9 Technical data

9.1 Dimensions

All measurements are stated in mm. All angles are stated in degrees.
All measurements and angles are stated without tolerances.
Invacare®reserves the right to change the stated measurements and
angles.
J
A (outer)
B (inner)
C (outer)
D
E
F
G
H
I
J
K

9.2 Weights

Max. patient weight
(provided that the weight of the mattress
and the accessories do not exceed 35 kg)
Max. safe working load
(patient + accessories)
Sonata standard bed, complete, excluding
accessories
Mattress support, head section four sections (2
pcs.)
16
A
K
890 – 1025 mm
2000 mm
2080 – 2130 mm
800 mm
230 mm
300 mm
610 mm
1230 mm
365 – 635 mm
0 – 72°
0 – 32°
135 kg
170 kg
15,5 kg
Mattress support, leg section four sections (2 pcs.)
Mattress support, four sections (1 pc)
Mattress support, four sections (1 pc)
Side rail Verso II (1 pc.)
Side rail Diana (1 pc.)
Side rail Flora (1 pc.)
Bed end Valeria (1 pc.)
Bed end Camila (1 pc.)
Bed end Sabrine (1 pc.)
Bed end Nadine (1 pc.)
Bed end Vibeke (1 pc.)
Lifting pole

9.3 Environmental conditions

Storage and
transportation
Temperature
-10°C to +50°C
Relative humidity
Atmospheric
pressure
Be aware that when a bed has been stored under low
temperatures, it must be adjusted to operating conditions
before use.

9.4 Electrical system

Voltage supply: Uin = 220 Volt, AC, 50 Hz (AC = Alternating
current)
Maximum current input: Iin max. = 1 A
Intermittent (periodic motor operation): Int = max. 10%, 2 min /
18 min
Degree of protection: IPX6*
Insulation class: Class II
Type B Applied Part
Applied Part complying with the specified requirements for
protection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which is
designed to come into physical contact with the patient or parts
that are likely to be brought into contact with the patient.)
Sound level: 45 to 50 dB (A)
* IPX6 classification means that the electrical system is protected
against water projected in powerful jets from any direction.
The bed is not provided with a mains switch, so the mains
plug is the only separation from the mains.
14.0 kg
28,5 kg
22,5 kg
7.0 kg
6.0 kg
8,5 kg
4.9 kg
11,0 kg
6,8 kg
5,0 kg
6,5 kg
5,0 kg
Operation
+5°C to +40°C
20% to 75%
800 hPa to 1060 hPa
1574228-D