Service - Mettler Toledo SevenCompact S210 Instrucciones De Manejo

Ocultar thumbs Ver también para SevenCompact S210:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Storing, viewing, printing and exporting data
Exporting settings to USB-stick
6. Screen settings
Screen Brightness
The screen brightness can be set from levels 1 to 16.
Screen Saver
The amount of time can be set which passes before the screen saver is activated:
5-99 minutes
If the meter is not operated during this time, the screen saver is activated. Press any key and the dis­
play is activated again, independent of the function of the key.
Screen Color
The display background color blue, grey, red or green can be selected.
Note
The display has a limited lifetime; therefore we recommend activating the screen saver or switch­
ing off the meter when not in operation.
If an instrument login password is set, this password is required after activating the display again.
5.7  Service
1. Software Update
In case a newer software version is available, the user can perform a software update via USB-stick,
by adhering to the following process:
1
Make sure that the firmware is in the root directory of the USB-stick and has a name
S<xxx>v<yyy>.bin, with <xxx> being the number of the instrument type and <yyy> being the
version number.
2
Connect the USB-stick to the instrument
3
Select the option "software update"
A message appears that the software update ist in progress
4
When the software update is completed you need to restart the instrument for the changes to
become effective.
Note
After the software update the instrument is switched back to factory settings. All unsaved data will
be lost and the PIN will be set back to "000000".
In case the USB-stick is removed during the update process or the power adapter is plugged out,
the instrument cannot be switched on anymore. Please contact METTLER TOLEDO service in that
case.
2. Export Settings to a USB-stick
With this feature the user can export the settings. These can for example be sent via e-mail to a cus­
tomer service representative in case difficulties occur, so that it is easier for the customer service repre­
sentative to provide support.
1
Insert the USB stick into the corresponding interface of the meter
The
icon appears on the display
2
Select the item and Export Settings to USB-Stick in the service menu to start the transfer
The settings will be in text (extension .txt) format. The instrument will create a new folder on the USB-
stick in which the name is the date in the international format, i.e. first year, then month and then the
day.
Setup
25

Publicidad

Tabla de contenido
loading