Timed Interval Readings; Temperature Settings; Measurement Limits - Mettler Toledo SevenCompact S210 Instrucciones De Manejo

Ocultar thumbs Ver también para SevenCompact S210:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1
Place sensor in the sample.
2
Press READ.
T appears on the display.
appears on the display to signalize measurement stability.
The measurement ends automatically when the set time period expires.
If READ is pressed before the signal is stable, the endpoint format changes to manual
Information on the display
The following symbols appear in the display, depending on the endpoint setting.
1
The actual endpoint format (last column) and not the preselected is stored with the data.
6.6  Timed Interval Readings
A reading is taken every time after a certain interval (1 – 2400 s) defined in the menu has elapsed.
When working in the timed-interval reading mode, the interval can be defined by entering the seconds.
The measurement series stops according to the selected endpoint format or manually by pressing
READ. When timed-interval reading is "on", the DL icon
The readings can be stored in the memory, transferred to the interface or both.
6.7  Temperature settings
Set MTC temperature
If the meter does not detect a temperature probe, MTC appears on the display. In this case the
sample temperature should be entered manually. An MTC value between -30 °C and 130 °C can
be entered.
Temperature unit
Select the temperature unit: °C or °F. The temperature value is automatically converted between
the two units.
The instrument works with NTC30 kΩ and Pt1000 temperature sensors. You can select between
automatic recognition or manual selection of the temperature sensor type. For temperatures below
100°C the instrument can reliably distinguish between NTC30 kΩ and Pt1000. At higher tempera­
tures, however, it is necessary to set the type of temperature sensor manually to avoid mistakes.
6.8  Measurement limits
The upper and lower limits for measurement data can be defined. If a limit is either not reached or
exceeded (in other words, less than or greater than a specific value), a warning is displayed on the
screen and may be accompanied by an acoustic signal. The message "outside limits" also appears
on the GLP printout.
30
Menus and settings
appears.
1
appears.
.
-