seT-up

seT-Up

Battery Connection

NOTE-
All models of LCSU 3 are shipped with the battery in place but not
connected. Follow the instructions below:
1�
Using a coin or straight-blade screwdriver rotate latch to unlocked
position (FIG� 1)�
2�
Remove door by pulling up on latch (FIG� 2)�
3�
Remove battery from compartment and plug connector into circuit
board (FIG� 3)�
4�
Replace battery and door; rotate latch to locked position�
5�
Fully charge battery for 5 hours before using (please see
Battery Charging)�
WARNING
Do not attempt to connect any type of suction tubing directly into the vacuum inlet port.
Only use with a Laerdal approved canister.
NOTE-
Always have a spare 300 ml canister within reach in case the canister in use is full or
the filter becomes wet.
88 00 50 Model (800 ml Configuration):
1�
Place canister into wire canister bracket; ensure inlet port marked <Patient> is accessible�
2�
Attach one end of the connection tubing to connection elbow (FIG� 4)�
3�
Insert unit connection elbow into side port (FIG� 5)�
4�
Attach other end of connection tubing to side of filter marked <Out>� NOTE-
Laerdal supplied replacement filters (FIG. 6).
5�
The 90° canister elbow connects to the clear side of filter marked <In> and to the top of
canister lid marked <Vacuum> (FIG 6)� NOTE-
canister.
6�
Connect the 1�8m (6') patient tubing to canister lid at inlet port labeled <Patient>�
7�
Ensure all connections are secure to prevent leakage in the canister/tubing system�
8�
Occlude the suction tube and set suction level according to local protocol before suctioning the
patient (FIG 7)� Be aware that it might be necessary to adjust the suction level during use�
88 00 60 Model (300 ml Configuration):
1�
This single-use sealed disposable canister includes an internal filter and fluid shut-off that
automatically stops suctioning when it becomes wet (FIG� 8)�
2�
Firmly attach canister by pushing connection fitting straight into the open side port of unit� (Both
the top connection fitting and the bottom support tab secure the canister) (FIG� 9)�
3�
Securely attach appropriate suction tip to the tubing�
4�
Occlude the suction tube and set suction level according to local protocol before suctioning the
patient (FIG 7)� Be aware that it might be necessary to adjust the suction level during use�
NOTE-
While suction unit will continue to operate if tipped on its side, canister capacity will be
reduced. Keep a spare 300 ml disposable canister within reach.
Figure 1
Use only
Verify clear side of filter marked <In> faces
5
5
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9